蛋白质吸附对医用金属材料体外腐蚀行为的影响
|
王鲁宁, 刘丽君, 岩雨, 杨坤, 陆黎立
|
Influences of Protein Adsorption on the in vitro Corrosion of Biomedical Metals
|
WANG Luning, LIU Lijun, YAN Yu, YANG Kun, LU Lili
|
|
表1 Mg及镁合金体内外测试中的腐蚀速率[3~7] (mm/a, unless otherwise indicated)
|
Table 1 The in vitro and in vivo corrosion rates of Mg and Mg alloys[3-7]
|
|
Material | In vitro | In vivo | Ref. |
---|
|
---|
| Solution | Weight loss | Hydrogen evolution | Volume reductions/ | | | | | | weight loss | Pure Mg | PBS | 0.28±0.07 | 0.19±0.02 | 0.15±0.03 | [3] | | HBSS | 0.72±0.31 | 0.57±0.07 | | | | DMEM | 1.07±0.10 | 0.57±0.07 | | | Mg-10Gd | PBS | 0.61±0.10 | 0.40±0.10 | 1.11±0.05 | [3] | | HBSS | 1.57±0.62 | 1.23±0.56 | | | | DMEM | 0.42±0.12 | 0.20±0.01 | | | Mg-2Ag | PBS | 16.71±3.06 | 15.12±2.97 | 0.13±0.04 | [3] | | HBSS | 5.41±0.71 | 3.48±0.36 | | | | DMEM | 2.21±0.26 | 0.68±0.04 | | | Pure Mg | DMEM | Static immersion 1.6±0.3 Aortic bioreactor: Lumen>4.0±0.8 Wall 1.9±0.5 | | Aortic lumen 1.1±0.3 | [4] | | | | Aortic wall 1.5±0.6 | | | | | | | | | | | | Mg-2Sr | HBSS | ~0.37 | ~0.32 | 1.37 | [5] | Mg-2Sr-Ca | HBSS | ~0.24 | ~0.21 | 1.10 | [5] | Mg-2Sr-Zn | HBSS | ~0.15 | ~0.11 | 0.85 | [5] | Pure Mg | HBSS | 17.2%±0.5% (14 d) | | 65.4%±1.5% (4 weeks) | [6] | Mg+Ca-Sr-P coating | HBSS | 4.8%±0.7% (14 d) | | 76.7%±1.9% (4 weeks) | [6] | Mg-30%Sc (mass fraction) | HBSS | | α+β phase 3.4±0.1 | β phase 0.06±0.01 | [7] | | | | Single phase 2.9±0.1 | | |
|
|
|